7-3 PCB Assembly Replacement

7-3
PCB Assembly Replacement
Procedures in this section are generic, and apply to many similar instruments. Photos and
Note
illustrations used are representative and may not match your instrument.
This section describes the removal and replacement of the SPA and Main/VNA boards which are attached to
each other and attached to the S412E Case.
1. Open the case as described in
2. Disconnect the Keypad PCB connector. For units with Ethernet hardware, unscrew and lift up the
Ethernet PCB to gain access to the Keypad PCB connector.
3. Disconnect the Fan Assembly connector, and the LCD connector.
4. Use a Phillips screwdriver to remove the 8 screws securing the Assemblies to the Case
Remove
2 screws
on side
Disconnect
Keypad PCB
Cable
Figure 7-4.
Removing the PCB Assemblies out of the Case
5. After the screws are removed the entire Assembly including the top connector panel will slide out of the
case.
6. Installation is the reverse of removal. During installation ensure the Keypad PCB cable along with all
other cables are properly seated at both ends. Also take care to properly fit the connector panel into the
grooves in the top of the case and confirm that none of the cables will be pinched when the back case is
replaced.
7-4
Section 7-2 "Opening the S412E
Remove 4 screws on the Motherboard
PN: 10580-00324 Rev. N
Chapter 7 — Assembly Replacement
Case".
(Figure
7-4).
Remove
2 screws
on side
Disconnect
Fan Cable
Disconnect
LCD Cable
S412E MM