7-7 Signal Generator Replacement
8. Remove the cable shown in
inserting the cable onto the new Signal Generator make sure the cable connector is parallel to the PCB
so the connector is plugged on correctly. The cable should not be removed or inserted at an angle.
Remove cable using
cable removal tool,
p/n 783-1399
Figure 7-11. Remove Cable Using Cable Removal Tool
9. Reverse the above steps to install the Signal Generator PCB.
7-10
Figure 7-11
using the cable removal tool, part number 783-1399. When
PN: 10580-00324 Rev. N
Chapter 7 — Assembly Replacement