Mobile Wimax Signal Analyzer Verification, Option 66 And 67 - Anritsu S412E Maintenance Manual

Lmr master vector network/spectrum analyzer
Hide thumbs Also See for S412E:
Table of Contents

Advertisement

5-5 Mobile WiMAX Signal Analyzer Verification, Option 66 and 67

5-5
Mobile WiMAX Signal Analyzer Verification, Option 66 and 67
The tests in this section verify the functionality of the Mobile WiMAX Signal Analyzer of the S412E.
There are tests for the following:
"Mobile WiMAX Channel Power Accuracy Tests (Option 66)"
"Mobile WiMAX Residual EVM and Frequency Error Tests (Option 67)" on page 5-13
Mobile WiMAX Channel Power Accuracy Tests (Option 66)
The tests in this section verify the function of the optional Mobile WiMAX Signal Analyzer in Model S412E
LMR Master.
Equipment Required
• Anritsu MG3700A Vector Signal Generator
• Anritsu ML2438A Power Meter
• Anritsu MA2482D Power Sensor
• Anritsu 34NN50A Adapter
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Aeroflex/Weinschel 1870A Power Splitter
• Aeroflex/Weinschel 44-10 10 dB Fixed Attenuator (2)
• Anritsu PN 2000-1627-R BNC Coaxial Cables (2)
• 10 MHz Reference Standard
Setup
TESTIME
MG3700
Vector Signal
Generator
250kHz-6GHz
From: 10 MHz Reference
To: S412E Ext Reference
Figure 5-2.
Mobile WiMAX Signal Analyzer Option Verification
5-10
10 MHz Reference
R
PLUS
R
RubiSource T&M
Sine wave Out
AUX
1 PPS OUT
1 PPS Sync
Failure
Oscillator Warmup
Oscillator Ready
Power
1 MHz
5 MHz
10 MH
5 MHz
10 MHz
1 PPS IN
Function
Ethernet
MG3700A Vector Signal Generator
PN: 10580-00324 Rev. N
ML2438A Power Meter
Cursor/Edit
RF Output
Control Input
Modulation Input
S412E
S412E LMR Master
Chapter 5 — Option Verification
MA2482D
Power Sensor
10 dB Attenuator
1870A
Power Splitter
10 dB Attenuator
N(m) to N(m) Adaptor
LMRMaster
ESC
Enter
Back
Shift
Mode
System
File
9
8
7
Limit
Trace
Measure
6
5
4
Sweep
Preset
Calibrate
3
1
2
.
+/-
0
Power
Charge
S412E MM

Advertisement

Table of Contents
loading

Table of Contents