Chapter 5 — Option Verification
13. Record the S412E Channel Power reading in the 797 MHz, –50 dBm row, Measured Channel Power
column of
Table
14. Calculate the Channel Power Error by subtracting the S412E "Channel Power" reading from the power
meter reading that was recorded in
column of
Table
15. Verify that the error is within specification.
Step 16
Step 20
through
16. Set the calibration factor frequency of the power sensor to 2150 MHz.
17. Set the MG3700A frequency to 2150 MHz. Press the Set key.
18. Change the S412E center frequency to 2150 MHz.
19. Measure the Channel Power for –10 dBm and –50 dBm and then record the measured result in the
Measured Channel Power column and the calculated error in the Error column of
20. Verify that the error is within specification.
LTE Frequency Error Verification (Option 542)
The tests in this section verify the function of the optional LTE Signal Analyzer in Model S412E LMR Master.
Procedure
1. Connect the MA2482D Power Sensor to the power meter and zero the sensor.
2. Set the Power Meter Measurement MODE to True RMS, set Averaging MODE to Moving, and set
Averaging NUMBER to 256.
3. Set the calibration factor frequency of the power sensor to 797 MHz.
4. Connect the equipment as shown in
The LTE pattern requires a Waveform Data license MX370108A that must be purchased.
Note
Both Set keys on the MG3700A perform the same function.
5. Set the MG3700A as follows:
a. Press the yellow Preset button (answer yes to the question).
b. Press the Set key
c. Press the (F1) soft key to select Load File to Memory.
d. Press the (F1) soft key again to select Select Package.
e. Using the Down Arrow key step through the selection list until LTE_DL_E-TM is highlighted.
f. Press the Set key.
g. Press the (F6) soft key Return.
h. Press the Set key. The Select Package list box will appear. Again select LTE_DL_E-TM and
press Set.
i. Another file list will appear. Select (highlight) E-TM_3-1_10M.
j. Press the Set key.
k. Press the MOD On/Off key and verify that the LED is On. Confirm that the "playing" indicator is
displaying the moving pattern.
l. Press the Frequency key, then enter 797 MHz.
m. Press the Level key, then enter 2 dBm.
n. Turn the output On.
S412E MM
A-30.
Step
A-30.
should be performed only for units with Option 6.
Figure
PN: 10580-00324 Rev. N
5-8 LTE Signal Analyzer Verification, Option 541 and 542
13. Record the result in the 797 MHz, –50 dBm row, Error
5-4.
Table
A-30.
5-21