Anritsu S412E Maintenance Manual page 57

Lmr master vector network/spectrum analyzer
Hide thumbs Also See for S412E:
Table of Contents

Advertisement

Chapter 5 — Option Verification
Mobile WiMAX Residual EVM and Frequency Error Tests (Option 67)
The tests in this section verify the function of the optional Mobile WiMAX Signal Analyzer in Model S412E
LMR Master.
Equipment Required
• Anritsu MG3700A Vector Signal Generator
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Anritsu PN 2000-1627-R BNC Coaxial Cable (2)
• 10 MHz Reference Standard
Setup
MG3700
Vector Signal
Generator
250kHz-6GHz
MG3700A Vector Signal Generator
Figure 5-3.
Mobile WiMAX Residual EVM and Frequency Error Test Setup
Procedure
1. Connect the equipment as shown in
Residual EVM and Frequency Error (10 MHz Bandwidth and 10 ms Frame Length)
2. Set the MG3700A as follows:
a. Press the yellow Preset button (answer yes to the question).
b. Press the Set key
c. Press the (F1) soft key to select Load File to Memory.
d. Press the (F1) soft key again to select Select Package.
e. Using the Down Arrow key step through the selection list until mWiMax is highlighted.
f. Press the Set key.
g. Press the (F6) soft key Return.
h. Press the Set key. The Select Package list box will appear. Again select mWiMax and Set.
i. Another file list will appear. Select (highlight) 10m1024g8_0_10_cap.
j. Press the Set key.
S412E MM
5-5 Mobile WiMAX Signal Analyzer Verification, Option 66 and 67
Function
Cursor/Edit
RF Output
Ethernet
Control Input
Modulation Input
Figure
5-3.
PN: 10580-00324 Rev. N
R
PLUS
R
TESTIME
RubiSource T&M
Sine wave Out
AUX
1 PPS OUT
1 PPS Sync
Failure
Oscillator Warmup
Oscillator Ready
Power
1 MHz
5 MHz
10 MH
5 MHz
10 MHz
1 PPS IN
S412E
LMRMaster
S412E LMR Master
ESC
Enter
Back
Shift
Mode
System
File
9
8
7
Limit
Trace
Measure
6
5
4
Sweep
Calibrate
Preset
3
2
1
.
+/-
0
Power
Charge
5-13

Advertisement

Table of Contents
loading

Table of Contents