Emerson NGA2000 Reference Manual page 84

Rosemount
Hide thumbs Also See for NGA2000:
Table of Contents

Advertisement

Reference Manual
748384-C
September 2003
DATA_POINT
Direction
Binding
Type
Value range
Units
Description
Usage
See also
DEBUG
Direction
Binding
Type
Value range
Units
Description
Usage
See also
DEBUGI
Direction
Binding
Type
Value range
Units
Description
Usage
See also
2-22
Analyzer Variables
Output
DATA_POINT
7-element floating point array
See description below
Depends on variables selected
Contains the data for the particular 15 minute average segment selected by
DATA_INDEX.
Set DATA_INDEX to the segment of stored data desired, then after about 1 second
this variable will be loaded with the data appropriate for that segment. If no data is
available for the particular segment, elements 0 and 5 will contain a large negative
number. The elements are as follows:
Primary variable (normally, but selected by TRENDCONT1)
Day of month
Hour
Minute
If > 9, calibration occurred; if 1 or 11, at least one invalid reading
Secondary variable (from TRENDCONT2)
The index where this data segment is stored, 0 - 95
Stored variables are selected by TRENDNAME, which shares data with
TRENDCONT to tell the analyzer which variables to store.
DATA_POINT
TRENDNAME
TRENDCONT
Output
None
7-element floating point array, 7-element integer array
Undetermined
None
Debug variables used in software development
Used only for software development, Undetermined contents
DEBUGI
Output
None
7-element floating point array, 7-element integer array
Undetermined
None
Debug variables used in software development
Used only for software development, Undetermined contents
DEBUG
NGA2000 Reference
Rosemount Analytical Inc.
A Division of Emerson Process Management