Emerson NGA2000 Reference Manual page 31

Rosemount
Hide thumbs Also See for NGA2000:
Table of Contents

Advertisement

NGA2000 Reference
presence of interfering compounds, such as
ammonia added to control NOx emissions.
You must remove these somehow.
After a span, the reading is very erratic...
You have spanned with a gas of much
lower concentration than you think, typically
zero gas or the measuring sample. Trou-
bleshoot the sample system until you are
sure that you are spanning off the real span
gas, and also enter into the analyzer menu
the real span gas concentrations. Enable
limit checking so that this does not happen
again.
The analyzer refuses to span...
If the analyzer considers that it has to
change its span factor by more than the al-
lowed amount, it will refuse to span, if the
calibration limit checking is enabled. Verify
that you have the correct span gas, and its
concentration is correctly entered into the
analyzer. If so, troubleshoot why the ana-
lyzer's span has changed so much.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Reference Manual
Possible cause are:
Contaminated sample cells in the NDIR
and CLD analyzers
Loss of ozone production in the CLD
Flame out in the FID
Incorrect flame position in the FID
Incorrect pressures and flows in all ana-
lyzers
Rapid temperature changes (greater
than 10°C per hour change)
Temperatures outside specification
Foreign substance contamination such
as water carry-over
Large ambient pressure changes
Sample leaks either within or outside
the analyzer
Exhaust back pressure changes
Incorrect linearizer coefficients
Span gas outside allowable range
748384-C
September 2003
Introduction
1-17

Advertisement

Table of Contents
loading

Table of Contents