Emerson NGA2000 Reference Manual page 91

Rosemount
Hide thumbs Also See for NGA2000:
Table of Contents

Advertisement

NGA2000 Reference
HIST2V_
Direction
Binding
Type
Value range
Units
Description
Usage
See also
I_TIME
Direction
Binding
Type
Value range
Units
Description
Usage
See also
IGNITECYCLES
Direction
Binding
Type
Value range
Units
Description
Usage
See also
IGNOVERIDE
Direction
Binding
Type
Value range
Units
Description
Usage
See also
Rosemount Analytical Inc.
A Division of Emerson Process Management
Output
None
2-element floating point array
Same as PRESS_IS for CLD
hPa
CLD only; values of PRESS-IS at manufacturing check out time
Provides historical data on how the analyzer was set up during final check out
PRESS_IS
Analyzer: bidirectional; I/0 module: input
TIME
7-element integer array
Appropriate date and time values
None
Periodic time update variable
Year
Mouth
Day
Hour
Minute
Second
Day of week -not implemented
Used to synchronize the internal clocks of analyzers and 1/0 modules with that of the
control module. The CM updates this variable every few seconds, and if it is updated
twice within 5 seconds, the analyzer or 1/0 module updates its internal clock to this
time.
Bidirectional
None
Integer
1-5
None
FID only; the number of ignition attempts allowed
Limits the number of ignition attempts during an automatic ignition cycle. If the flame
has not ignited by this time, the algorithm will give up.
Bidirectional
None
Integer
2-10
none
FID only; the duration of the glow plug current flow
The glow plug requires a current flow for a certain period in order for it to warm up
enough to ignite the fuel. This variable sets that time. It is normally 3 seconds
IGNITECYCLES
Reference Manual
748384-C
September 2003
Analyzer Variables
2-29