Emerson NGA2000 Reference Manual page 83

Rosemount
Hide thumbs Also See for NGA2000:
Table of Contents

Advertisement

NGA2000 Reference
CURRENTZERO
Direction
Binding
Type
Value range
Units
Description
Usage
See also
CURRENTZFAC
Direction
Binding
Type
Value range
Units
Description
Usage
See also
DATA_INDEX
Direction
Binding
Type
Value range
Units
Description
Usage
See also
Rosemount Analytical Inc.
A Division of Emerson Process Management
Output
None
Floating point
Same as ZEROGAS
ppm
The value of the zero gas for the current range.
Used to show the current range's zero gas on menus, rather than the entire array of
span gases as would be shown by ZEROGAS. The analyzer copies the value of the
appropriate element of ZEROGAS into this variable whenever CRANGE is changed.
ZEROGAS
CRANGE
Output
None
Floating point
Same as ZERO
ppm
The value of the zero factor for the current range.
Used to show the current range's zero factor on menus, rather than the entire array
of zero factors as would be shown by ZERO. The analyzer copies the value of the
appropriate element of ZERO into this variable whenever CRANGE is changed.
ZERO
CRANGE
CURRENTSFAC
Input
DATA_INDEX
2-element integer array
Element 1: 0 - 96; element 2: 0 - 59, or up to 65535
None
Index into the locally stored data communicated in DATA_P0INT
The two elements have alternative meanings: if the second element has a value less
than or equal to 59, the second element refers to the time in minutes, and the first to
the time in hours - the time referring to the desired data point. If the value of the
second element is greater than 59, the first element then is taken to refer to the
index of the data point.
DATA_POINT
Reference Manual
748384-C
September 2003
Analyzer Variables
2-21