NGA2000 Reference
STCONT
Direction
Binding
Type
Value range
Units
Description
Usage
See also
SVCONT
Direction
Binding
Type
Value range
Units
Description
Usage
See also
SVFLL
Direction
Binding
Type
Value range
Units
Description
Usage
See also
SVFUL
Direction
Binding
Type
Value range
Units
Description
Usage
See also
Rosemount Analytical Inc.
A Division of Emerson Process Management
Analyzer output and 1/0 module output
STCONT
8-element integer array
Bit field variables; odd elements are 16 bits, even elements use only the last 4 bits
None
Each bit of the first element shows the state of the corresponding alarm named in
ST1NAME; the odd elements show the first 16 elements of ST1NAME, the even show
the last four. Similarly for the remaining pairs with ST2NAME etc.
Used to show the state of the specific analyzer alarms listed in ST1NAME, ST2NAME
and so on.
ST1NAME
Analyzer and 1/0 module: output
None
4-element enumerated array
A list of variable names, each element having the same list.
None
Contains the variable names for the control module to use in the secondary variable
display.
The variable SVNAME shares data with this variable. When a selection is made in
SVNAME, it causes the control module to use the contents of the equivalent
enumeration in SVCONT as the value to be displayed on the auxiliary line of the single
analyzer display screen. Element 1 is displayed on line 1, and so on. The description in
SVNAME is used as the text description on the variable line on the same screen.
SVNAME
Bidirectional
None
Floating point
0 - about 200
cc/min
The sample bypass flow rate lower limit
Used in the alarm system to provide the lower limit for the flow alarm. When the flow-
rate given in FLOW_IS drops below this value, the variable VALIDITY is set to invalid.
SVFUL
FLOW_IS
Bidirectional
None
Floating point
600-1200
cc/min
The sample bypass flow rate upper limit
Used in the alarm system to provide the upper limit for the flow alarm.
SVFLL
FLOW_IS
Reference Manual
748384-C
September 2003
Analyzer Variables
2-55