1-26 SYNCHRONIZING AUTOCAL MODULES

Autocalibration modules can be synchronized
by assigning them master, slave or independent
status. A Master will force any slaves in the
same system to start their calibration sequence
when the master does. A Slave will not perform
a calibration unless the Master tell it to. An In-
dependent module will ignore a master, and
simply work on its own internal timing.
Rosemount Analytical Inc.
A Division of Emerson Process Management
a.

Basic Instructions

Enter Expert controls and setup
Enter Auxiliary module setup...
Select the module.
Enter General parameters...
Select This module's control status:
Edit this to MASTER, SLAVE or INDE-
PENDENT.
Reference Manual
748384-C
September 2003