The Service functions menu
Reagent status of the retort
Display of printer status
ASP6025 Tissue Processor
(continued)
• The ASP6025 software continuously records which reagent was the last
reagent in the retort. The last reagent in the retort is displayed next to the
button. This way, it is ensured that incompatible reagents do not come into
contact with one another.
• If a filling process was canceled, it is possible that two reagents are displayed
Xylene
here.
• If the user knows that there actually was a different reagent in the retort than
the one that is displayed, this can be corrected by selecting the LAST RETORT
REAGENT button.
• After the button has been pressed, first a warning message appears which
must be confirmed. Then, the current reagent list is displayed and it is thus
possible to change the reagent status of the retort by selecting the correct
reagent.
Fig. 107
The status of the PDF printer is displayed here. It cannot be changed.
"Printer ready" means that a PDF file can be generated and saved after a
program has run, if this was specified in the run options (see also
Printer
Fig. 94).
ready
Fig. 108
Caution!
The reagent status of the retort may not be changed unless it
is known with certainty that the displayed reagent is not cor-
rect. If incompatible reagents come into contact with one an-
other, this can result in a malfunction of the instrument or the
impairment of a process.
5.
Operation
Chapter 5.1.2,
83