The RMS system – adjusting warning threshold values
Tables with warning threshold values
ASP6025 Tissue Processor
Xylene protocols
Reagent
Formalin
70 % ethanol
85 %/90 % ethanol
1 100 % ethanol
1 xylene
1 paraffin
Cleaning xylene
Cleaning alcohol
Cleaning water
for a low throughput of approx. 100 cassettes daily
1)
Xylene-free protocols
Reagent
Formalin
70 % ethanol
85 % ethanol
1 (ethanol/isopropanol) 80/20
1 isopropanol 100 %
ParaLast
1 paraffin
Cleaning xylene
Cleaning alcohol
Cleaning water
for a low throughput of approx. 100 cassettes daily
1)
(continued)
recommended replacement
after 600 cassettes or 2 cycles
after 600 cassettes or 2 cycles
after 1500 cassettes or 5 cycles
after 1500 cassettes or 5 cycles
after 1500 cassettes or 5 cycles
after 1800 cassettes or 6 cycles
after 6 cycles
after 6 cycles
after 6 cycles
recommended replacement
after 600 cassettes or 2 cycles
after 600 cassettes or 2 cycles
after 1500 cassettes or 5 cycles
after 1500 cassettes or 5 cycles
after 1500 cassettes or 5 cycles
after 2400 cassettes or 8 cycles
after 2400 cassettes or 8 cycles
after 6 cycles
after 6 cycles
after 6 cycles