Maintenance task
Paraffin station (paraffin)
Clean the system bottles
Clean the RTU-bottles
Check to see if the system bottle connections are perma-
nently screwed into place
Clean the outer surface of the ASP6025
Check the filter in the paraffin station and clean it if
necessary
Check and clean the lids of the paraffin baths
Clean the paraffin drip tray (retort)
Drain the condensate bottle
Carry out a SMART clean
Replace the active carbon filter (every 30-45 days)
Carry out a full cleaning with detergents
Clean system bottles
ASP6025 Tissue Processor
6.
After each
Daily
run
X
Cleaning and Maintenance
Weekly
Monthly
X
X
X
X
X
X
X
X
X
X
X
X
165