The System monitor menu
Active carbon filter
ASP6025 Tissue Processor
(continued)
Fig. 104
Also in Supervisor mode, only the runtime of the active carbon filter
can be reset in the SYSTEM MONITOR menu
If the ACTIVE CARBON FILTER button is touched, a field appears
which displays the programmed warning and alarm threshold values.
Even in supervisor mode, neither of these values can be changed.
• If the stored warning value is exceeded, a warning message
regarding the age of the carbon filter is output.
This warning serves as a reminder to replace the carbon filter. You
can continue using the instrument without any limitations.
• If the set alarm threshold value is exceeded, a message is output
on the screen to inform the user that the carbon filter age has
been exceeded.
• When this message appears, at the latest, the active carbon filter
must be replaced in order to ensure that it functions and works
efficiently.