Emptying the condensate bottle
Replacing the active carbon filter
13
When inserting the new filter, make sure that
it is pushed all the way into the compartment
and that the flap is closed.
ASP6025 Tissue Processor
14
Fig. 231
Filter label
Fig. 233
6.
Cleaning and Maintenance
• The blue condensate bottle must be taken out, emptied and
reinserted at regular intervals (weekly). This should be done
sooner if a prompt to that effect is issued by the software
(Fig.
231).
Dispose of waste solvents with care according
to local regulations and the waste manage-
ment policy of the company or institution.
• The life of the active carbon filter will depend on the reagent
types used and how frequently they are used in the instru-
ment.
• The filter (13) should be replaced after 45 days at the latest
(Fig.
233).
• We recommend a SMART clean prior to changing the active
carbon filter.
• After replacing the filter, open the SYSTEM MONITOR menu
to set the filter replacement date to the current date. See also
Chapter 5.1.4.
Fig. 232
163