Instrument Setup
Pull-out drawer with RTU-bottles
10
64a
Spigot closed
50
(continued)
• Seven RTU-bottles (10) are supplied for the drawer with the instrument
- 6x for tissue infiltration,
- 1x for the cleaning process (extended cleaning program by detergent).
Each container has a screw plug.
• Each container has a max. capacity of 5 l.
Only RTU-bottles from Leica are approved for this instrument. However, if
d ifferent bottles are used, the customer must INDEPENDENTLY check that they
are suitable for such use.
(For information about temperature and dimension requirements
Fig. 46
Warning message.)
Spigot open
The bottom right of the drawer has a spigot (Fig.
draining spilled or overflowing reagents neatly. Then, the
drawer should be cleaned.
Important!
Always keep the spigot closed
tail, item 64a).
The spigot may be opened for cleaning pur-
poses only
erwise flammable reagents could end up on
the heater of the paraffin baths and ignite.
Fig. 47
64b
Fig. 47a detail
(Fig.
46)
see page
(Fig. 47a de-
(Fig. 47a
detail, item 64b), as oth-
Instructions for Use V 2.4 RevL - 12/2015
51-
47), for