Draining/filling reagents
Select station
+
• The entire process is displayed graphically on the screen
(SMART SCREEN view)
ASP6025 Tissue Processor
(continued)
Press button
Fig. 164
Fig. 165
(Fig.
166).
Fig. 167
•
On the screen (Fig.
162), select the system bottle or RTU-
bottle that is to be drained and then select the receptacle
(external station).
Then press STATION DRAIN.
• A prompt appears for carrying out the necessary preparations
(Fig.
165). This must be confirmed with YES.
• The reagent is pumped out of the RTU-bottle into the retort.
• Once this task has been completed, the retort is drained into
the external station (79 in
• A message window on the screen shows that the process has
been completed
(Fig.
167). To continue working, confirm this
with OK.
• If the same reagent will not be used in the next operating
step, the retort may have to be cleaned with a cleaning
program.