5.
Operation
Draining/filling reagents
Select station
+
120
(continued)
Press button
Fig. 168
Fig. 169
Fig. 170
Remote filling
Remote filling of system bottles takes place in exactly the same
way as remote draining of RTU-bottles, but in the reverse order
of course.
• On the screen, select an empty system bottle (e.g. S6) and
the external receptacle (Remote Station,
• Then press the STATION REFILL button.
• Now the reagent with which the highlighted bottle is to
be filled (of course, the reagent contained in the external
station) must be selected
• After you have selected a reagent, a screen message appears
prompting you to make the necessary preparations.
• When everything is finished, press YES to start filling
(Fig.
170).
Instructions for Use V 2.4 RevL - 12/2015
Fig.
168).
(Fig.
169).