Leica asp6025 Instructions For Use Manual page 135

Vacuum tissue processor
Table of Contents

Advertisement

Starting a program
(continued)
RUN OPTIONS
This function
(Fig.
198) shows the options with which the
current program was started.
The options set apply for ALL programs!
All run options can be enabled or disabled according to requirements while the program is running.
ASP6025 Tissue Processor
as the end time, an entry window
can select the end time.
To ensure that the program ends on the
specified date and time, the ASP6025 extends
the duration of the programmed delay step
correspondingly.
When working with fixed end times, you
Fig. 197
should always check that the instrument time
and date are current.
• PRE-TEST activated:
After the program is started, the retort is filled and drained
with the first reagent in the infiltration program to test
whether all lines and valves are unobstructed.
• PRINTER, if activated:
At the end of the program, a complete run log covering the
entire program execution is stored in the form of a PDF file.
Fig. 198
5.
If you start a pro-
gram for which TIME
REQUEST was selected
(Fig.
197) opens in which you
Pressing the RUN OPTIONS button opens
the window with the same name.
Operation
135