Leica asp6025 Instructions For Use Manual page 173

Vacuum tissue processor
Table of Contents

Advertisement

• If no infiltration program is running when a power failure is detected, no further actions / backup measures are taken -
the instrument shuts down automatically.
When a program is running:
• The program that is running is stopped, regardless of the current status of processing.
To conserve the battery, all heaters, the agitator and the monitor are switched off directly.
Current step for power failure
Formalin (fastening)
Ethanol (Dehydration, diluted)
Ethanol (absolute)
Xylene (intermedium)
Isopropanol (intermedium)
Paraffin
ParaLast
ASP6025 Tissue Processor
• If a reagent is being used for which no safe reagent is in the
• If a safe reagent is in the retort, the reagent can be kept in
• If no safe reagent is in the retort, depending on the cur-
Fig. 245
Protection in the safe reagent
Formalin
Formalin
Formalin
Xylene (heating off)
Formalin
Paraffin (heating off)
ParaLast
instrument when the program starts, a warning message
245). However, the program launch is NOT
appears
(Fig.
aborted.
the event of power failure, though the retort may have to be
filled to a high enough level so that no tissue sample is dry.
This operation is automatically controlled by the instrument
and cannot be influenced by the operator.
rent status of processing the program that is in progress, a
"decision" is made about how to proceed.
7.
Troubleshooting
173