4.
Instrument Setup
The retort
(continued)
Fill level sensors
40.
4
40.
3
40.
2
40.
1
Level sensors
(at left in Fig.
Selection
buttons
Filling the retort up to sensor 40.
Filling the retort up to sensor 40.
Filling the retort up to sensor 40.
42
40
31)
1
2
3
Fig. 31
Insert the sieve (42) into the drain opening on
the floor of the retort.
42
It is imperative to insert the mag-
netic stirrer before an infiltration
program is started!
Attach the magnetic stirrer (41) to the axis on
the floor of the retort with the little hole facing
41
Every time a program is started (except in CONCENTRATION
mode, see
Chapter
5.1.2), a message is first displayed in which
the start must be confirmed.
Depending on how many baskets the retort is loaded with, the
fill volume can be selected in the dialog box by pressing the
corresponding button
(Fig.
31).
• For this purpose, there are three level sensors (40.
(Fig.
31) located behind the cover (40) in the rear cover of
the retort.
• The uppermost sensor (40.
overfilled or overflowing.
Instructions for Use V 2.4 RevL - 12/2015
upwards.
Fig. 30
- 40.
1
) prevents the retort from being
4
)
3