Leica asp6025 Instructions For Use Manual page 122

Vacuum tissue processor
Table of Contents

Advertisement

5.
Operation
15.
1
122
Manually changing the reagents
35
15
• Remove the system bottle (15) from the reagent cabinet and
65
• Drain the used reagent into a bulk container and then refill
Fig. 174
• Place the filled system bottle back in the reagent cabinet.
• When inserting the system bottle, always make sure that the
• Then, in the SETTINGS window press the REAGENT STATUS
• Select the corresponding line in the table and set the data
Fig. 175
Caution!
Never replace reagents or fill empty system
bottles or RTU-bottles while a process is run-
ning.
This can result in severe damage to the instru-
ment.
unscrew the container cover (15.
the system bottle.
The fill levels for filling with 3.8 l and 5.0 l are embossed
on the front of each container.
Use a funnel for clean filling.
retaining ring (35) is securely tightened and the connection
spout (65) clicks into the back panel correctly.
button and call up the table of the same name
to zero using the DATA TO ZERO button
Manual filling has been described previously
with reference to the system bottles.
Follow the exact same procedure for the RTU-
bottles from the drawer.
Instructions for Use V 2.4 RevL - 12/2015
).
1
(Fig.
175).
(Fig.
175).