Leica asp6025 Instructions For Use Manual page 16

Vacuum tissue processor
Table of Contents

Advertisement

2.
Safety
Warnings – Maintenance and cleaning
• Do not use solvents that contain acetone for cleaning. Liquid must not get into the instrument while
operating or cleaning it.
• When using cleaners, comply with the safety instructions of the manufacturer and the safety regulations of
your laboratory!
• Check the condensate container at least once a week and, if necessary, drain it.
• System bottles and Leica RTU-bottles must not be cleaned in a dishwasher - the containers are NOT dishwasher-
proof.
Material safety data sheets can be obtained from the supplier of the chemicals.
Alternatively, they can be downloaded from the following website:
http://www.msdsonline.com
Problem solution for hazard warning
Warning!
Misuse of the emergency unlock function and improper opening of the retort
Injuries to hands, arms and head when opening the retort and/or injuries due to spraying reagents!
• Only use the mechanical emergency unlock function if the standard emergency unlock
not work and you absolutely must get to the specimens.
• Wear special protective clothing, gloves and safety goggles.
• Grasp the bolt handle of the retort firmly in your hand.
• Ensure that no part of the body is located in the opening area of the retort lid.
• Since the retort may be under pressure, open the bolt handle very carefully and slowly.
16
(see Chap.
7.4.1) did
Instructions for Use V 2.4 RevL - 12/2015