Power Supply PCBA Removal

Required Parts

• P19909-XX Service Assembly, Power Supply PCBA
• V19909-XX Vet Service Assembly, Power Supply PCBA
Power Supply PCBA Removal
1 Remove Seven long screws and two short screws securing the digital shield lid. Please note the location
of the 2 short screws as seen in
hole should be empty and covered with kapton tape. Do not place a screw in this location. Remove lid
and set aside.
Long Screws
(7x)
Short Screws
(2x)
Screw missing
intentionally.
Should be covered
by kapton tape
Figure 6.16 Power Supply Shield
2 Gently lift the Power Supply PCBA and Power Supply Shield away from the Main PCBA as shown in
Figure 6.17
(The Power Supply PCBA connector is in the upper left corner. Removing the Power Supply
PCBA without also removing the shield is difficult).
3 Once disconnected from the Main PCBA, separate the Power Supply PCBA from the Power Supply
shield.

Power Supply PCBA Replacement

1 Place the Power Supply PCBA in the Power Supply shield and install on Main PCBA.
2 Install the Power Supply Digital Shield Lid.
3 Install the seven long screws and two short screws. Torque to 5.5 in-lbs.
50
Chapter 6: Replacement Procedures
Figure 6.16
. Also note the location marked with the black box. The screw
Power Supply
Shield