Chapter 5: Troubleshooting
This chapter contains information to help you correct problems with system operation.
Note: If the system requires repair, certain steps must be taken to remove patient data from the
system prior to return to FUJIFILM SonoSite. To accomplish this, a Power-0 Reset must be
performed whenever possible (certain conditions may prevent this, such as a system that fails to
power on).
Power-0 Reset formats the system SD memory.
Warning: This will erase all patient data saved on the system, but will not delete an already
established partition for the transducer database. Patient images should be exported or archived
before proceeding.
The procedure for performing the Power-0 Reset is as follows:
Start with the system powered off.
1) Press and hold the "M" key.
2) Press and release the system On/Off button while continuing to hold down the "M" key.
3)When the system emits a high-low sound, release the "M" key. This sound will occur after 5-10
seconds and indicates that the system is reset. The reset will be complete about 1 minute after the
system has booted up, when the system emits a low-high sound.
Power-1 Reset causes the system settings to revert back to factory defaults. This includes the
following:
Sets system date/time to January 1, 2003, 02:00.
Sets IP address to 169.254.254.254. DICOM or SiteLink network configurations already saved to
the system will still present, but must be selected again in the Connectivity screen.
Resets the Audio, Battery settings.
The procedure for performing the Power-1 Reset is as follows:
Start with the system powered off.
1) Press and hold the "2D" key.
2) Press and release the system On/Off button while continuing to hold down the "2D" key.
3) Release the when the system emits a high-low then low-high sound. This sound will occur after
5-10 seconds and indicates that the system is reset.

System and Subsystem Diagnosis

This section covers basic diagnostic and troubleshooting procedures you may follow if the system
does not operate properly. To diagnose system failures, consult the referenced diagnostic figures
that follow or FUJIFILM SonoSite Technical Support.
Table 5.1: Troubleshooting Subassemblies and Diagnostic Figures
Subassemblies
DICOM
User Interface
Battery
Diagnostic Figures or Table
Table 5.2
Table 5.3
Table 5.4
Chapter 5: Troubleshooting
33