Specifications - Sartorius M3P Installation And Operating Instructions Manual

Table of Contents

Advertisement

Specifications

Model
Electronic weighing range
Maximum loading capacity
Readability
Tare range (by subtraction)
Standard deviation
(corrected for drift, typical*)
Max. linearity
Stabilization time (typical)
Display update
Adaption to ambient conditions
and application requirements
Stability range
Ambient temperature range
Sensitivity drift within 290 ... 300 K
Deviation of the readout when the
balance is tilted 1:1000
Pan size
Balance housing (W x D x H)
Weighing chamber (W x D x H)
Net weight, approx.
Power requirements
(voltage + frequency: 50 – 60 Hz))
Allowable voltage fluctuation
Power consumption
Interface
* for samples or objects that are not or only slightly electro statically charged
M3P
mg
500/1,000/1,510 (Poly Range)
mg
approx. 3,000
mg
0,001/0,002/0,005 (Poly Range)
mg
approx. –3,000
mg
≤±0,001/0,002/0,003
≤±0,002/0,004/0,005
mg
s
10
s
0,1 – 0,8 (selectable)
By selection of 1 of 4 optimized filter levels
d
0,25 ... 64 (selectable)
o
C
+15 to +30 (59
-6
≤±5∙10
/K
mg
≤±0,03
mm
Ø 22 (about 1 inch)
mm
219 x 291 x 137,5
inches
8.6 x 11.5 x 5.4
mm
54 x 49,5 x 53,5
inches
2.2 x 1.9 x 2.1
kg
5,2 (approx. 11.4 lbs.)
115 or 230 V depending on the power supply used
-20% ... +15%
VA
max. 13 (typical)
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
transmission rate 150 ... 9,600 Baud
o
o
F to 86
F)
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3p-000v001

Table of Contents