Specifications
Model
Electronic weighing range
Maximum loading capacity
Readability
Tare range (by subtraction)
Standard deviation
(corrected for drift, typical*)
Max. linearity
Stabilization time (typical)
Display update
Adaption to ambient conditions
and application requirements
Stability range
Ambient temperature range
Sensitivity drift within 290 ... 300 K
Deviation of the readout when the
balance is tilted 1:1000
Pan size
Balance housing (W x D x H)
Weighing chamber (Dia. x H)
Net weight, approx.
Power requirements
(voltage + frequency: 50 – 60 Hz))
Allowable voltage fluctuation
Power consumption
Interface
* for samples or objects that are not or only slightly electro statically charged
28
M3P-000V001
mg
500/1,000/1,510 (Poly Range)
mg
approx. 3,000
mg
0,001/0,002/0,005 (Poly Range)
mg
approx. –3,000
mg
≤±0,002/0,003/0,004
mg
≤±0,002/0,004/0,005
s
15
s
0,1 – 0,8 (selectable)
By selection of 1 of 4 optimized filter levels
d
0,25 ... 64 (selectable)
o
C
+15 to +30 (59
-6
/K
≤±5∙10
mg
≤±0,03
mm
105/73/22 (~4"/3"/1")
mm
219 x 291 x 98
inches
8.6 x 11.5 x 3.9
Diameter depends on the choice of filter weighing pan: for
105 mm pan: 117 mm (4.6")
for 73 and 22 mm pans: 85 mm (3.3")
Effective height for all versions: 11 mm (.43")
kg
5,4 (approx. 11.9 lbs.)
115 or 230 V depending on the power supply used
-20% ... +15%
VA
max. 13 (typical)
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
transmission rate 150 ... 9,600 Baud
o
o
F to 86
F)