3.10 STARTING CURRENT OR OVERCURRENT (True Inrush™)
MEASUREMENT
To measure a starting current or overcurrent, proceed as follows:
1.
Set the switch to
measured. Press the
2.
Perform a long press on the
displayed, along with the triggering threshold. The clamp then awaits
detection of the True InRush™ current. "------" is displayed and the A
symbol flashes.
3.
After detection and acquisition for 100 ms, the RMS value of the True-
Inrush™ current is displayed. Pressing the
PEAK+/PEAK- values subsequently.
4.
A long press on the
switch will exit the True InRush™ mode.
NOTE: The triggering threshold in A is 10A if the initial current is zero (starting of
installation). For an established current (overload in an installation) see §3.4.3.
3.11

FREQUENCY MEASUREMENT (HZ)

The frequency measurement is available in V and A for AC measurements. The
measurement is based on a count of zero crossings (positive-going edges).

3.11.1 Frequency Measurement (V)

To measure the frequency in voltage, proceed as follows:
1.
Set the switch to
2.
Select AC by pressing the
is reached.
3.
Connect the black lead to the COM terminal and the red lead to the "+"
terminal.
4.
Connect the test probes or the alligator clips to the circuit to be
measured.
, then clamp the jaws around the conductor to be
(yellow) button to select AC measurement.
button or a change of function on the rotary
and press the
(yellow) button until the desired choice
25
button. The InRh symbol is
button will display the
button. The Hz symbol is displayed.