4.4 REAR PANEL SW1 SETUP SWITCH

The SW1 Setup switch (see Fig.4-3) is a 9-position DIP switch that allows the user to choose the
following:
- Internal or remote programming for Output Voltage and Current Limit.
- Remote voltage or resistive programming of Output Voltage and Output Current limit.
- Select range of remote voltage and resistive programming.
- Select range of Output Voltage and Output Current monitoring.
- Select the Remote Shut-Off control logic.
- Select between RS232 or RS485 communication interface.
- Enable or disable the rear panel Enable/Disable control (dry contact).
Fig.4-3: SW1 setup DIP switch
4.4.1 SW1 position functions
Refer to Table 4-3 for description of SW1 position functions. The factory default setting is Down
for all positions.
Table 4-3: SW1 positions functions
Position
Function
Output Voltage
SW1-1
Remote analog programming
Output Current Limit
SW1-2
Remote analog programming
Programming range select
SW1-3
(Remote voltage/resistive)
Output Voltage and
SW1-4
Current Monitoring range
SW1-5
Shut Off logic select
SW1-6
RS232/485 select
Output Voltage
Resistive programming
SW1-7
SW1-8
Output Current Limit
Resistive programming
Enable/Disable control
SW1-9
4.4.2 Resetting the SW1 switch
Before making any changes to the SW1 switch setting, disable the power supply output by
pressing the front panel OUT button. Ensure that the output voltage falls to zero and OUT LED is
off, then use any small flat-bladed screwdriver to change the SW1 switch setting.
9
8
7
6
5
4
3
2
1
Down (Factory default)
Output Voltage
Programmed by
Front Panel
Output Current Limit
Programmed by
Front Panel
0-5V / (0-5Kohm)
0-5V
On: High (2-15V) or Open
Off: Low (0-0.6V) or Short
RS232 interface
Output Voltage
Programmed by
External Voltage
Output Current Limit
Programmed by
External Voltage
Rear panel
Enable/Disable control is
not active
32
32
Up
Output Voltage
Programmed by remote analog
External Voltage or Extenal Resistor
Output Current Limit
Programmed by remote analog
External Voltage or External Resistor
0-10V / (0-10Kohm)
0-10V
On: Low (0-0.6V) or Short
Off: High (2-15V) or Open
RS485 interface
Output Voltage
Programmed by
External Resistor
Output Current Limit
Programmed by
External Resistor
Rear panel
Enable/Disable control is
active