Adjustment
resistor
(in ohm)
0
0.39
0.82
1.2
1.6
2
2.7
3
3.3
3.9
4.7
5.6
42

Selecting the Adjustment Resistor

Adjusting Off-center Load with 1000 Ohms Initial Resistance

Test weight
Test weight
50kg
100kg
0.00 - 0.02
0.00 - 0.05
0.02 - 0.04
0.05 - 0.07
0.04 - 0.06
0.08 - 0.11
0.06 - 0.07
0.12 - 0.15
0.08 - 0.09
0.16 - 0.19
0.1 - 0.13
0.20 - 0.26
0.13 - 0.14
0.27 - 0.29
0.15 - 0.16
0.30 - 0.32
0.16 - 0.19
0.33 - 0.38
0.19 - 0.23
0.39 - 0.46
0.23 - 0.27
0.47 - 0.55
0.20 - 0.33
0.56 - 0.66
Adjusting the Weighing Platform
Test weight
Test weight
200kg
Off-center (in kg)
0.0 - 0.1
0.0 - 0.1
0.1 - 0.1
0.1 - 0.4
0.1 - 0.2
0.4 - 0.5
0.2 - 0.3
0.6 - 0.8
0.3 - 0.4
0.8 - 1.0
0.4 - 0.5
1.0 - 1.3
0.5 - 0.6
1.3 - 1.5
0.6 - 0.7
1.5 - 1.6
0.7 - 0.8
1.6 - 1.9
0.8 - 0.9
1.9 - 2.3
0.9 - 1.1
2.3 - 2.7
1.1 - 1.3
2.8 - 3.1
Important: These are calculated (estimated) values
Test weight
500kg
1000kg
0.0 - 0.3
0.3 - 0.7
0.8 - 1.1
1.2 - 1.5
1.6 - 1.9
2.0 - 2.6
2.7 - 2.9
3.0 - 3.2
3.3 - 3.8
3.9 - 4.6
4.7 - 5.5
5.6 - 6.6