Cable from the Load Cell to the Junction Box 2 (Terminals)
Color Coding
See page 33 for the colors of the wires; the casing is gray or black.
Function:
Screen (Shield)
BBR_POS (Bridge supply voltage +)
BR_NEG (Bridge supply voltage -)
SENSE_POS (Sense +)
SENSE_NEG (Sense -)
OUT_POS (Measuring voltage positive)
OUT_NEG (Measuring voltage negative)
OUT_POS (Measuring voltage positive)
To connect the cable, the wires must be connected to the inner terminals (marked
in the illustration on page 20/22).
Note:
The color coding of the connecting cable (A/D converter / indicator to
junction box) might be different when connecting a non-Sartorius
platform to the Midrics indicator.
Adjusting the Weighing Platform
Terminal:
Supply +
Supply -
Sense +
Sense -
Signal +
Signal -
Signal +