Display
Inf 07
Inf 08
Inf 09
Inf 22
Inf 23
Inf 29
Inf 30-31
Inf 71
Inf 72
Inf 73
Cause
Function not allowed in scales
verified for use in legal metrology
The load on the scale is too heavy to zero
the readout
Taring is not possible when the gross weight
is a minus value
Error in storing reference value, load is too
light
Error in initializing an application
Minimum load not reached
Indicator is in XBPI mode
Cannot store the current weight value
(e.g., if control limits are too low or too high)
Cannot store the current weight value
(e.g., the transaction counter has reached
its limit)
Data not found or unreadable
Solution
Contact your local Service Center
for details on changing settings
Check whether "Tare/zero at power on" is set
(1.12)
Zero the scale
Put a heavier weight on the scale
Contact your local Service Center
Define a lower value for the minimum load
(in the Application settings, menu item 3.6)
Change menu setting or perform "Close" function.
Check the interface and cable.
Contact your local Service Center.
None
None
Contact your local Service Center