8.9

Replacing the restraining belt

Requirements:
● 10 mm socket wrench
● 4 mm Allen key
Uninstalling the restraining belt:
● Loosen the screw (3) and the corresponding nut (not shown in the illustration) with a
4 mm Allen key and a 10 mm socket wrench.
● Remove the nut including the washer.
● Remove the screw together with the restraining belt and the washer (2) and also the washer
behind it.
Note
Another nut is used between the washers as a spacer, so that the belt bracket can move freely.
● Replace the restraining belt (1).
Installing the restraining belt:
● Assemble the parts in reverse order.