8.2.3

Replacing the G-Trac module

Requirements:
● 10 mm socket wrench
● Side cutting pliers
● Cable ties
Uninstalling the G-Trac module:
Switch the electronics of the wheelchair OFF.
Remove the front, rear and right lateral covers, as described in chapter 8.1.1, 8.1.2 and 8.1.3.
The G-Trac module is located on the spring-seat of the right-hand spring.
Trace the cable of the G-Trac module (1) to the electronic module and disconnect it there.
Pull the cable out of the wheelchair. If necessary remove any attached cable ties.
Use the 10 mm socket wrench to loosen and remove the self-securing nuts (3) including the
washers.
Pull the G-Trac module (2) off of the carriage bolt (1) which is inserted in the spring seat.