Table 6-2 First Current Output Calibration Procedure - Honeywell UDC3500 Product Manual

Universal digital controller
Hide thumbs Also See for UDC3500:
Table of Contents

Advertisement

Procedure
The procedure for calibrating the First Current Output is listed in Table 6-2. Make sure
that LOCK in the Tuning Set Up group is set to NONE. (See Subsection 3.4 – Loop 1
Tuning Set Up Group.)

Table 6-2 First Current Output Calibration Procedure

Step
Operation
Enter Calibration
1
Mode
2
Calibrate 0 %
Calibrate 100 %
3
4
Exit the
Calibration Mode
10/05
Press
Upper Display = CALIB
Setup
Setup
Lower Display = CURRENT
until you see
You will see:
Func
Loop 1/2
Upper Display =
Lower Display = ZERO VAL
Until the desired 0 % output is read on the milliammeter,
or
use the values shown below depending on the action of
your controller. Normally, this will be the setting that
produces 4 mA.
This stores the 0 % value and
Func
you will see:
Loop 1/2
Upper Display =
Lower Display = SPAN VAL
Until the desired 100 % output is read on the milliammeter,
or
use the values shown below depending on the action of
your controller. Normally, this will be the setting that
produces 20 mA.
The controller stores the span value.
Func
Loop 1/2
To exit the calibration mode.
Lower
Lower
Lower
Display
Display
Display
UDC3500 Universal Digital Controller Product Manual
Result
A Value
A Value
Output Calibration
277