NORMA 4000/5000
Operators Manual
In the Timing & Sync Setup menu, define the following settings:
Column
Tavg[s]
Src
Level
Slope
Filter
SyOut

Set Average Time

The average time is a multiple of the period of the voltage of current source. The settings
are automatically adjusted during measuring. For example: the average time is set to 19
ms; at a frequency of 50 Hz, it is automatically adjusted to 1 period, that is, 20 ms.
Short average times are useful, if you wish to analyze individual periods,
measuring even minute interferences. With long average times, for example
300 ms at 50 Hz, short-term interferences are not shown.
Value in column Tavg[s] is highlighted.
7-10
Settings
15 ms... 3600 s
U1 / I1 ... U6 / I6
ext
Off
−150 % ... +150 %
↑ or ↓
10 kHz
1 kHz
100 Hz
off
On
Off
1. Press Enter.
A window with a numerical entry field is displayed.
2. Enter the first digit of the average time and confirm by pressing Enter.
Repeat the above step for the other digits.
Description
Minimum average time (in seconds)
Synchronization source
Fixed average time
Trigger level (in % of measuring range)
Slope direction
Synchronization filter (filter is not in signal path)
Signal output enabled
Signal output disabled (at Sync Ext output)
Note
esn021.eps