Integration Function Configuration; Integration Setup - Fluke NORMA 4000 Operator's Manual

Power analyzer
Hide thumbs Also See for NORMA 4000:
Table of Contents

Advertisement

The correct parameter settings are shown on the type plate of the shunt or
probe.
You can:
Enter the transducer ratio (U over I) at the external current meters and instruct the
device to calculate the final scale factor.
Or, enter the scale factor at the current transducer so that the final scale factor can
be calculated.
The parameters of the formula must be entered as follows:
Scale factor x transducer ratio, whereby:
Scale factor: generally "1.0000" (one).
Transducer ratio: current (in ampere) to voltage (in volt).
If you select Set all to apply the configuration to all channels, only the scale
factor is transferred. If shunt values U/I are entered, the scale factor is
always 1, and Set all is not available. If probes are used, it is generally
easier to enter the transducer ratio, and Set all is thus not recommended.
1. Move the cursor to the value in column Scale and press Enter.
A dialog window showing the scale formula is displayed.
2. Select a value for each parameter and press Enter to confirm.
The settings are shown in column Scale.
3. Save the configuration settings by pressing the SAVE function key.

Integration Function Configuration

This configuration controls key parameters required for the calculation of integrated
values over time.
You can select up to six different integration parameter (values) out of a list.
Active power P1 to P3 and the sum power are preselected.

Integration Setup

When the Power Analyzer is switched on; the start screen is displayed.
1. Press function key WAV.
The integration symbol is displayed in the assignment bar for function keys.
Note
Note
Note
Configuration

Integration Function Configuration

esn024.tif
7