7.2.3.3. Average time dependent signal drop (averaging mode)
The signal drop follows an averaging time function. The average time is the amount of time the
output signal needs to reach 90% magnitude compared to a perpendicular drop. This parameter
is set by means of the programming command <AA>.
Temp

Figure 33: Average Time Dependent Signal Drop (averaging mode)

7.3. Advanced Peak Hold

This function searches the sensor signal for a local peak and writes this value to the output
until a new local peak is found. Before the algorithm restarts searching for a local peak, the
object temperature has to drop below a predefined threshold. If the object temperature raises
above the held value which has been written to the output so far, the output signal follows the
object temperature again. If the algorithm detects a local peak while the object temperature is
currently below the predefined threshold the output signal jumps to the new maximum
temperature of this local peak. Once the actual temperature has passed a peak above a
certain magnitude, a new local peak is found. This magnitude is called hysteresis.
The threshold is set by means of the programming command <C>, for hysteresis use the
command <XY>.
Temp

7.4. Valley Hold

This function works similar to the peak hold function, except it will search the signal for a
minimum. The output signal follows the object temperature until a minimum is reached. The
output will hold the minimum temperature value for the selected duration of the valley hold
time. Once the hold time is expired, the valley hold function will reset and the output will
resume tracking the object temperature until a new valley is reached. The range for the valley
hold time can be set from 0.1 to 300.0 seconds, whereas just 0.1 - 299.9 seconds will be
interpreted as valley hold duration. A value of 300.0 seconds indicates that valley hold post
processing depends on an external trigger signal. A low level input (GND) at external input
Innovative High Temperature Infrared Pyrometers
hysteresis

Figure 34: Advanced Peak Hold

output temperature
object temperature
Time
output temperature
object temperature
Time
39