Change View Mode

After having selected a channel and the relevant measured values, you have the option to
change to different view modes where the parameters are shown in the form of numerical
values, vector graphs or oscilloscope graphs.

Numerical Display

For details regarding the numerical display of measured values, refer to the section,
"Measure Voltage, Current and Power," in Chapter 8.

Vector Graphs

Up to six signals of the H01 fundamentals can be viewed as vector graphs.
The vector graphs show voltage and current with amplitude and phase shift, and allow for
the fast assessment of signals and detection of errors in the connections.

View Vector Graphs

Display
φ1 ... φ3
φU1
φU2
φU3
scale
8-16
1. Press measuring keys Σ or 1...n and WAV to call up the desired values, that is,
values measured at WAV power.
2. Press measuring key Vector graphs.
The measured values are shown in the form of vector graphs.
Phase angle between U and I
... reference point (always = 0)
Angle between U2 and U1
Angle between U3 and U1
Range (reference value for the diameter of the outer circle)
3. To view a different channel or different measured values in vector graph form,
press measuring keys Σ or 1...n and WAV.