Switch Current Input to External Input (BNC)

If you want to use an external shunt or probe you have to change the current input from
direct measurement to the BNC input. This has to be done in the Current Channel Setup
menu.
This procedure consists of the following steps:

Switch Current Input

First field in column Range is highlighted.

Configure Auto-Range Selection

First field in column Auto is highlighted.
Auto range is now enabled.

Configure Scale

If you intend to use a shunt or a probe, you must adjust the scale for the output of the
measuring signal.
7-18
Call up Current Channel Setup
Switch input
Configure input range

Configure scale

1. Press Enter, select a voltage range (such as, 3 V) and press Enter.
2. If you wish to configure all three current channels in this way press Set All.
3. Save the configuration settings by pressing the SAVE function key.
The current input is now changed to the external BNC input.
1. Press Enter, select ON and confirm with Enter.
2. If you wish to configure all three current channels in this way press Set All.
3. Save the configuration settings by pressing the SAVE function key.