Current Channel Setup - Fluke NORMA 4000 Operator's Manual

Power analyzer
Hide thumbs Also See for NORMA 4000:
Table of Contents

Advertisement

NORMA 4000/5000
Operators Manual
The configuration procedure consists of the following steps:

Current Channel Setup

Move the cursor to the status display of current channel I1 and press Enter.
The Current Channel Setup menu is displayed. The first field in column Auto of line I1 is
highlighted.
In the Current Channel Setup menu, define the following settings:
Column
Ch
Auto
Range
Scale
Coup
Filter
7-14
Call up Current Channel Setup
Configure input range
Configure scale
Configure coupling
Configure anti-aliasing filter
Call up Voltage Channel Setup
Settings
I1 ... I6
ON
OFF
30 mA ...10 A
30 mV ... 10 V
Scale factor and A/V ratio
AC
DC
ON
OFF
Description
Select input (channel)
Automatic range adjustment
activated
... deactivated
Measuring range (in ampere or volt)
Scale for external probes/converters
Coupling
Filter activated
... deactivated