Chapter 3—Setting Treatment Parameters

New Treatment Key

When the 2008K is first turned on in preparation for daily operation (after a long power down),
all treatment parameters revert to their default settings. This can also be accomplished by
pressing the New Treatment (New TX) key. The New TX key allows the operator to reset patient
treatment parameters to their default settings without interrupting the power to the machine. This
must be done in preparing the 2008K for all subsequent treatments after the first of the day.
After pressing the New TX key or after a long power down:
All treatment data (blood pressure, Kt/V) are deleted. The treatment summary information
is moved to the previous record in the Tx Summary screen.
The RTD counter is reset to zero
All heparin treatment parameters are reset to zero
SVS profile is reset to None
UF treatment parameters are reset as follows:
UF profile is reset to None
UF Removed is reset to zero
UF Goal = 3000
UF Time = 3:00
UF Rate = 1000
To activate the New Treatment Option:
1) Press
New Tx
Figure 18 – New Tx key
48
2) Press
CONFIRM
2008K Operator's Manual rev. K