158
Pressure Test Failed
The pressure test section (PHT) of
the automated Test Sequence has
failed.
Priming
The operator has pressed the Prime
key and initiated the priming function.
Put Connectors in Port
The connectors must be in the
machine ports in order to start a
Rinse program
Put Lines On Shunt
An action has been initiated that
requires the dialysate lines to be on
the shunt door.
Put Red Con In Chemical
This is a cleaning/disinfectant
program prompt to the operator.
RAM Battery Failure
Electronic self-test failure.
RAM Code Corrupted
Electronic self-test failure.
RAM Code Corrupted 2
Repeated electronic self-test failure.
Recirc Error 1
This Message may occur during the
Chemical Rinse program
Recirc Interrupted
The Recirculate program has been
interrupted by an alarm condition.
Recirculating
Recirculation is in progress.
Troubleshooting
Alarm
Reset the alarm and repeat the test. If the failure message is
repeated on retest, take the machine out of service and alert a
qualified technician.
Advisory
Advisory only. No action is required.
Advisory
Connect the red (acid/acetate) and/or blue (bicarbonate) connectors
to the appropriate rinse ports.
Warning
To proceed with the selected operation, place dialysate lines on the
shunt.
Advisory
Remove the red wand/connector from the machine and place it into
the chemical/disinfectant bottle.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Warning
Refer machine to a qualified technician.
Warning
1) Inspect the blood pump condition. Correct if required.
2) Reset the alarm and turn UF back on, if applicable.
3) If the alarm does not clear, take the machine out of service and
alert a qualified technician.
Advisory
Advisory only. No action is required.
2008K Operator's Manual rev. K