134
BIC PUMP ALARM
This is a pump failure warning.
BIC PUMP ALWAYS
This is a pump failure warning.
EOS
BIC PUMP NO EOS
This is a pump failure warning.
BIC Conn Out Of Port
The blue bicarbonate wand/connector
is out of its port.
Bicarbonate has been set
The operator has attempted to set a
to max.
Bicarbonate level higher than allowed.
Bicarbonate has been set
The operator has attempted to set a
to min.
Bicarbonate level lower than allowed.
Blood flow unstable
When attempting to start an OLC test,
certain conditions are necessary,
including stable blood flow rate.
Blood Leak not Calib
The blood leak detector is not in
calibration.
Troubleshooting
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, take the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, take the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, take the machine out of service.
Warning
Connect blue (bicarbonate) wand/connectors into the blue rinse port.
Verify the concentrate selection.
Dialog
The Bicarbonate will be set to the highest allowed bicarbonate level.
Message
Confirm if value is acceptable or enter new value.
Dialog
The Bicarbonate will be set to the lowest allowed bicarbonate level.
Message
Confirm if value is acceptable or enter new value.
Advisory
Wait a minute or so and start the OLC test again.
Alarm
Take the machine out of service and alert a qualified technician.
2008K Operator's Manual rev. K