Fresenius Medical Care 2008K Operator's Manual page 108

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 4—Monitoring the Treatment
3. Discard the bloodlines and transducer protectors according to facility policy
4. Clean or disinfect machine according to routine cleaning and maintenance procedures
described in "Disinfection and Maintenance," on page 109.
If Reuse is not practiced
To remove the fluid in the dialysate compartment:
1. Open shunt interlock door
2. Return blue dialysate connector to shunt interlock
3. Reposition the dialyzer so that the red, outlet port is at the bottom
4. Close the shunt interlock door. Message "Emptying" will be displayed.
5. Drain the dialysate compartment. The dialyzer is empty as soon as there is air in the
outlet line or an "Emptying stopped" message appears.
6. Open the shunt interlock door, remove dialysate outlet line from the dialyzer and place it
on the shunt.
7. Discard the bloodlines, transducer protectors, and dialyzer according to facility policy.
8. Insert the concentrate wands into their proper rinse ports. The "Select Program" screen
appears on the display screen.
9. Clean or disinfect the exterior of the machine according to routine cleaning and
maintenance procedures described in "Disinfection and Maintenance," on page 109.
Removing Bloodlines from the Machine
The arterial and venous ends of the bloodline should be clamped to avoid spillage before
attempting to remove the lines from the system.
Caution: Do not forcefully pull the lines from the machine. Damage to the machine or its
components may result.
To remove the bloodline from the blood pump, open the door and align the rotor by pressing
and holding the Start key until the pump stops. Press the latch below the rotor to release the
left (incoming) side of the pump segment. Pull the first couple of inches of the pump
segment out of the pump. Then, while keeping firm tension outward on the left (incoming)
side of the bloodline, press and hold the Start key a second time and the pump segment will
be released from the pump head.
Be sure to open the door to the optical detector before pulling the line from the venous clamp
and optical detector assembly.
108
2008K Operator's Manual rev. K