Fresenius Medical Care 2008K Operator's Manual page 144

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

144
E.98 (Arterial or SN
Error erasing Flash ROM while in
Blood Pump Message)
service mode
E.99 (Arterial or SN
Transmit error during Flash update
Blood Pump Message)
while in service mode
Eeprom already used.
Advisory message when uploading
Power off and replace
hardware key option
eeprom
EEPROM Missing or
During startup, the machine cannot
Reading Error
properly read the EEPROM memory
chip
Emptying
The blue dialysate line connector is
on the shunt with door closed, the red
dialysate line connector remains on
the dialyzer in order to drain the
dialysate compartment.
Emptying Stopped
When air is sensed, emptying will
stop.
Emptying too long
The dialyzer empty program has
exceeded its maximum limit.
Error Reading Flash
Electronic Self Test
Fail*9 Volt Battery
9V Power Failure Battery test has
failed.
Fail*Actuator Arterial High
Arterial Pressure test has failed.
Fail*Actuator Arterial Low
Conductivity test has failed.
Fail* Actuator
Conductivity High
Fail* Actuator
Conductivity Low
Troubleshooting
Alarm
If this message occurs in dialysis mode, take the blood pump module
out of service and alert a qualified technician.
Alarm
If this message occurs in dialysis mode, take the blood pump module
out of service and alert a qualified technician.
Advisory
Put in a new hardware key or calibration EEPROM in IC 20 and
power up.
Opening
Turn the machine off and try to power up again. If the message
Screen
repeats, refer to a qualified technician.
Message
Warning
If this message occurs when the dialyzer is not being emptied, take
the machine out of service and alert a qualified technician.
Warning
Connect the red dialyzer line to the shunt. If the warning is repeated,
take the machine out of service.
Alarm
If blood is not sensed, return the dialysate lines to the shunt and
close the shunt door to terminate the program. If the machine was in
dialysis (blood sensed), turn machine power off and back on to clear
the program.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Test
Replace Battery
Message
Test
Rerun test, if failure repeats, remove from service.
Message
Rerun test, if failure repeats, remove from service.
Test
Message
2008K Operator's Manual rev. K

Hide quick links: