Using a Central Venous Catheter

Shock Hazard: Ensure that no conductive electrical devices connected to or near the patient
have leakage currents above the maximum CF applied parts limit of 10 μA DC and 50 μA DC
in a single fault condition. Failure to follow these precautions may result in serious injury or
death.
Maintenance
Assembly, installation, adjustment, or repair is to be performed only by persons authorized by
the facility medical director or by Fresenius USA.
Questions?
For further information regarding the operation, repair, parts, or maintenance of the Fresenius
2008K hemodialysis machine, please contact:
Fresenius USA
2637 Shadelands Drive
Walnut Creek, CA 94598
16
catheter
Warning:
patient is
electrically
connected to
ground via blood
and dialysate
Do not allow
contact with
additional
electrically
conductive
devices with
leakage
currents
greater than
10 µA DC
or greater than
50 µA DC
"single fault"
(800) 227-2572
(925) 295-0200
2008K Operator's Manual rev. K