Message
Purpose of Message
Recirculating Done
A prompt to the operator that the
recirculation process is done.
Recirculating Stopped
Recirculation has been stopped
because blood is sensed or the
dialysate lines are on shunt.
Release CONFIRM to
This message is a prompt for the
stop
operator to release the CONFIRM key
to stop priming the heparin syringe.
Rel. Blood Volume Low
The BVM module has reported a
relative blood volume below the lower
limit.
Reset CRRT Time
CRRT time has reached 99 hours and
59 minutes
Reset to adjust TMP
The TMP exceeded the hard alarm
limits. The operator is given the
option to relieve the pressure to bring
the TMP within limits.
Reset Treatment?
The New Tx key has been pressed
Resetting, Try Again
Blood pressure module resetting
Retry > Press XXX
The cuff pressure is too low to
measure the blood pressure. The cuff
pressure is XXX mmHg.
159
Advisory
Warning
Message
Warning
Warning
Warning! Adjusting the
TMP repeatedly will
decrease the UF removed
from the patient.
Advisory
Warning
Message
Troubleshooting
Type
Action Required
Press RESET to clear the advisory message.
Inspect the configuration of the dialysate lines and extracorporeal
blood circuit. Correct any irregularities. If the message is not cleared,
take the machine out of service and alert a qualified technician.
Dialog
Release the CONFIRM key.
Alarm
Press RESET to clear the message. Evaluate the fluid status of the
patient.
Reset UF Removed to 0
Press the RESET key to reset the TMP alarm limits. Press and hold
the Override key to re-center the limits.
Warning! Rising TMP may indicate a leak in the balancing
system and should be investigated.
To reset treatment parameters for a new treatment, press the
CONFIRM key. To cancel, press the Escape key.
Wait until blood pressure module completes resetting and retry blood
pressure reading.
Dialog
No action necessary
2008K Operator's Manual rev. K