OPERATOR'S INSTRUCTIONS
Fresenius Medical Care North America
920 Winter St.
Waltham, MA 02451
Manufactured by:
Fresenius USA, Inc.
4040 Nelson Avenue
Concord, CA 94520
Copyright  1998 - 2014 Fresenius Medical Care North America
P/N 490039 Rev. C
Page 1 of 3

Advertisement

loading

Summary of Contents for Fresenius Medical Care DIASAFE 2008 H

  • Page 1 Fresenius Medical Care North America 920 Winter St. Waltham, MA 02451 Manufactured by: Fresenius USA, Inc. 4040 Nelson Avenue Concord, CA 94520 Copyright  1998 - 2014 Fresenius Medical Care North America P/N 490039 Rev. C Page 1 of 3...
  • Page 2 OPERATOR'S INSTRUCTIONS  FOR THE 2008H & 2008K WITH DIASAFE FILTER Warning: With highly permeable dialyzers, ultrafiltration must be carried out only with volume- controlled ultrafiltration. Warning: For the safe execution of hemodiafiltration and dialysis with highly permeable dialyzers it is important that the balancing system of the 2008H or 2008K Dialysis Machine is operating properly.
  • Page 3 Manual Pressure Holding Test Place the dialysate lines in the shunt with dialysate flow at 500 ml/min. Set UF rate to 500 ml/hr. Allow the UF pressure to rise until a TMP of 300 (260-340) mmHg is established. Stop the UF pump by turning the UF OFF. Allow the TMP to stabilize. The TMP may change by no more than 60 mmHg within 3 minutes.