OPERATOR'S INSTRUCTIONS FOR THE 2008H & 2008K WITH DIASAFE FILTER Warning: With highly permeable dialyzers, ultrafiltration must be carried out only with volume- controlled ultrafiltration. Warning: For the safe execution of hemodiafiltration and dialysis with highly permeable dialyzers it is important that the balancing system of the 2008H or 2008K Dialysis Machine is operating properly.
Page 3
Manual Pressure Holding Test Place the dialysate lines in the shunt with dialysate flow at 500 ml/min. Set UF rate to 500 ml/hr. Allow the UF pressure to rise until a TMP of 300 (260-340) mmHg is established. Stop the UF pump by turning the UF OFF. Allow the TMP to stabilize. The TMP may change by no more than 60 mmHg within 3 minutes.